Speaking Valves

Passy-Muir Speaking Valves

In the previous section on communication options, we reviewed different types of ways an individual with tracheostomy may speak such as mouthing words, writing, communication boards, electrolarynx, assistive technology, leak speech, finger occlusion, talking tracheostomy tubes, speaking valves and flap valves.  

Here we will review the different types of speaking valves on the market for patients with tracheostomy tubes and the differences between them.  Each individual resident is different and one valve may be more appropriate than another based on patient comfort, benefits and goals. 

Passy-Muir Tracheostomy and Ventilator Swallowing and Speaking Valves (PSMV)

Passy-Muir Speaking Valves
Images curtesy of Passy-Muir, Inc. Irvine, CA.

The Passy-Muir Tracheostomy & Ventilator Swallowing & Speaking Valve is often referred to the shortened name of Passy-Muir Valve or by the acronym PMV or PMSV.  PMV is the term for prolonged mechanical ventilation and therefore if an acronym is used in a medical setting PMSV is preferred to reduce confusion.  

The Passy-Muir Valve was invented by David Muir.  David had muscular dystrophy and required a tracheostomy and mechanical ventilation.  He was unable to speak.  In a time of frustration, he invented the Passy-Muir Valve, which has provided thousands of voices to individuals with tracheostomy and mechanical ventilation all over the world.  

.  

Passy-Muir Valve Design

The Passy-Muir Valve is the only bias-closed position no -leak speaking valve on the market.  It is important to understand the design in order to understand the unique benefits that result.

Click the video below to watch how the Passy-Muir Valve works.

Differences in Passy-Muir Valves

Passy-Muir has developed several models that are available.  The PMV 005 (white), PMV 007 (aqua), PMV 2000 (clear) and the PMV 2001 (purple) which all can be used on both individuals spontaneously breathing with tracheostomy or those with mechanical ventilation.  

 

All models fit with any tracheostomy tube size, neonatal, pediatric or adult.  The only requirement is that the tube must have the universal 15mm hub.  

Tracheostomy Tube parts

The PMV™ 005 is the original version and has mostly been replaced with the PMV™ 2000.  Both these valves are designed for individuals in a home environment as they are lower profile.  The PMV™ 007 and 2001 are for use in an institutional setting to alert staff that the speaking valve is in place or for color preference in a home environment.  The PMV 007 can be used with disposable ventilator tubing for ventilator application.  The PMV™ 005, 2000 and 2001 require rubber non-disposable tubing to fit securely in-line for ventilator application.  

All Passy-Muir Valves have the same bias-closed position design.  The Passy-Muir Valve has independently been found to produce the best speech quality, as reported by listeners and participants (Leder, 1994). 

The closed position “no leak” design of the Passy-Muir Valves promotes the establishment of a more normal “closed respiratory system.  These valves are designed in a biased-closed position and offer less work of breathing. There is no air leakage out of the Passy-Muir Valve during expiration.  Instead, all of the exhaled air is directed around the tracheostomy tube, up through the vocal cords, and then out through the oral and nasal cavities. Since the Passy-Muir Valve remains closed except when the patient inspires, a column of air is trapped within the valve and tracheostomy tube which acts as a buffer to resist the movement of secretions up the tracheostomy tube and onto the valve. 

Due to the unique closed-position, the Passy-Muir Valve offers benefits that other speaking valves do not provide. 

Visit  Passy-Muir.com for more information.

Open Position Speaking Valves

Open position speaking valves are one way valves that open to allow inspiration to occur through the tracheostomy tube, but require reverse airflow from exhalation to close the membrane of the valve. This air leakage during exhalation can permit secretions to travel up into the tracheostomy tube and onto the valve.  The inability of an open position valve to completely seal negates any benefit other than to restore voice for the tracheostomized patient and may provide some ability to cough and expel secretions through the upper airway.  Examples of open position valves include Shikani Speaking Valves, Shiley PhonateTracoe Phon Assist, and Montgomery Speaking Valve.  

Shikani Speaking Valves™ (SSV)

Shikani speaking valve and comparisons
Clear Shikani Speaking Valve
Shikani Speaking Valve
Blue Shikani Speaking Valve

Shikani Speaking Valves are another option for individuals with tracheostomy to use to produce speech.  The SSV uses a unique ball design and a strategically placed chamber.  There is a slight tilted ramp that guides the ball toward the front of the chamber with expiration.  The ball moves towards the back of the chamber due to the act of inspiration and then the ball is redirected to the  front of the chamber with expiration. The expired air moves the ball to allow speech to travel through the upper airway, taking the path of least resistance.  

The patient can vary the position of the valve (valve “up” or valve “down”) by rotating it 180°, which allows the ball to be seated in the proximal part of the chamber. 

Valve Up Diagram

Valve up (12 o'clock)

This is the valve "up" position with the notch and green line are facing up. Upon resting breathing, this allows airflow in and out of the tracheostomy as if there were no valve in place at all. When the patient speaks, the ball is pushed up to close the frontal opening, thus redirecting airflow through the upper airway for speech.

Valve Down Diagram 1

Valve down (6 o'clock)

This is the valve in the "down' position with the notch and green line facing down. When the valve is in this position, the ball automatically moves forward, closing the frontal opening. Inhalation directs the ball posteriorly to allow airflow through the tracheostomy tube and into the lungs. Exhalation pushes the ball back to the closed position, to redirect airflow through the upper airway.

Some patients may be more comfortable when the valve is in an alternative position, such as 3 o’clock or 9 o’clock.  

Shikani speaking valves are considered open position valves, resulting in some air leaking out of the valve on expiration.  The benefit of air leakage is that there is less resistance, making it easier to breathe on exhalation.  The drawback is that there will be a loss of positive airway pressure so that the cough reflex will not be as strong and subglottic pressure would escape through the tracheostomy tube, which is important for swallowing, secretion management and decannulation.  An open position design may benefit individuals who do not tolerate the biased closed-position design of the Passy-Muir Valve.  However, it is recommended to first determine the reason if the patient is having difficulty exhaling through the upper airway.  An ENT referral may help to determine the etiology.  

Since there is leakage of airflow during exhalation when using the Shikani Speaking Valve, it is possible to simultaneously use an HME device.  The Shikani HME™ functions as an artificial nose and is designed to restore some of the essential respiratory functions which are lost when the breathed air is directed away from the nose. In particular, the Shikani HME™ filters airborne environmental particles from air inhaled by the patient with a tracheostomy; it retains moisture and heat from air that is exhaled out of the lungs; and upon re-inhalation, it filters, humidifies, and warms air returned to the lungs.

Shikani Speaking Valve and Shikani HME Combination
Shikani HME™

 Visit The Airway Company for more information about Shikani Speaking Valves.

Shiley™ Phonate Valve

Shiley Speaking Valve and comparison to other valves
Medtronic 2019. All rights reserved. Permission to use from Medtronics.

The Shiley Phonate Valve is an open position, one way valve designed for use on alert, awake tracheostomy patients who breathe spontaneously.  This valve is not compatible for individuals on mechanical ventilation. Compared to finger occlusion, it reduces hygiene and infection concerns.  One model is available with a built in nipple port for the addition of supplemental oxygen.

The Shiley Phonate Valve opens on inhalation to allow airflow to the lungs.  Exhalation is required to force the membrane of the valve to the closed position.  Therefore, some air escapes out the tracheostomy tube at the end of inspiration and during exhalation which may allow secretions to build inside the tracheostomy tube.  A flip top lid was designed to allow for secretion removal and cleaning of the Shiley Valve.  

Visit Medline for more product information. 

Tracoe phon assist I

The Tracoe phon assist I is an open position speaking valves for individuals with tracheostomy, breathing spontaneously.  It is not compatible for patients on mechanical ventilation.  The valve has a silicone membrane which closes during exhalation at a fast occlusion time (25 ms), to redirect airflow through the upper airway.  There are two side openings to adjust the air flow resistance individually.  When the valve is turned to allow for an opening in the side, airflow will escape both through the upper airway as well as through the tracheostomy tube on exhalation.  Therefore by increasing the side openings, there will be less resistance and easy breathing with the valve in place.  The more closed the valve, the more airflow is directed through the upper airway.  There is an oxygen supply port to provide supplemental oxygen. 

Tracoe phon assist comes in both bright orange for improved visibility as well as in white. 

See Tracoe for more product information. 

Montgomery® Speaking Valves

Montgomery® Tracheostomy Speaking Valve

The Montgomery®  Tracheostomy Speaking Valve is an open position, one way valve designed for use on awake tracheostomy patients who breathe spontaneously. The valve is not to be used for individuals on mechanical ventilation.  The valve provides one-way airflow using a thin silicone diaphragm that opens on inspiration and closes on expiration.  An exclusive feature of this valve is a “cough-release” mechanism, eliminating valve or tube dislodgment as a result of coughing or excessive airway pressure.

Montgomery® Ventrach™ Speaking Valve

Montgomery® Ventrach™ Speaking Valve is an open position, one way valve designed for individuals on mechanical ventilation. It provides airflow through the membrane of the speaking valve on inspiration and closes on expiration to redirect airflow through the vocal folds and upper airway.  Significant air loss has been demonstrated during exhalation, when the valve should be closed (Zajak et al, 1999). 

Speech Therapy with a Speaking Valve

Once a speaking valve is in place, speech therapy can focus on increasing valve wear time, improving voicing, improving breathing and speech coordination.  

With children, use of a speaking valve may be a frustrating time due to the changes in airflow.  Music can be a less stressful approach to therapy.  

Finding Your Voice- Songs for Kids with Tracheostomy is the perfect adjunct for therapy.  There are 15 songs plus 3 instrumentals to help ease children to use the upper airway.  Our Animal Faces and Tracheostomy Tube with Speaking Valve are great to use either with the songs or on their own.  They can be used to help children imitate.  All animal face cut outs are made on laminated 8.5×11 poster board.  

 

Summary

speaking valve comparisons chart

Speaking valves allow patients with tracheostomy the opportunity to produce speech naturally.  Exhaled air is redirected around the tracheostomy tube and deflated cuff (if present), through the vocal folds and upper airway.  The cuff of the tracheostomy tube must always be completely deflated (if present) or the patient will be unable to breathe.  The Passy-Muir Valve is the only bias-closed position valve which provides additional benefits.  The Passy-Muir Valve and Montgomery Ventrach may be used in-line with mechanical ventilation.  

See ventilator application of speaking valves for more information about in-line speaking valve use with mechanical ventilation. 

Related Webinars:

Resources

Leder SB. Perceptual rankings of speech quality produced with one-way tracheostomy speaking valves. J Speech Hear Res 1994;37:1308-12

Zajak, D, Fornataro-Clerici, L, Shoop, R. Aerodysnamic Characteristics of speaking valves. J Speech Hear Research 1999; 42; 92-100. 

Subscription Form

Subscribe for free to get our latest tracheostomy and mechanical ventilation posts, new products, and information about our new membership (coming soon!)

Related Articles

COVID-19 Tracheostomy and Mechanical Ventilation

The 2019 novel corona virus COVID-19 pandemic has resulted in an increase in patients intubated and use of mechanical ventilation. The United States and globally, we are likely to see an increase in tracheostomy as well, as patients may have difficulty weaning and require longer periods of time on a vent. COVID-19 also has implications for healthcare workers, as there are shortages with workers becoming ill from the virus. Infection control is paramount in controlling the outbreak and protecting patients, healthcare workers and the community